Manual installation of FraunDiff 1.0 under Windows: At installation, the following procedure must be followed: 1. Make sure that appropriate DAC drivers (Computerboards, Measurement computing, e.g. CIO-DAS08-AOH) are installed and properly configured. 2. Create the FraunDiff folder and copy the following files from the installation media: FraunDiff\ - folder FraunDiff\Fraundiff.exe - file FraunDiff\Fraundiff.hlp - file FraunDiff\Fraundiff.cnt - file FraunDiff\Lesmeg.txt - file FraunDiff\Data - folder FraunDiff\Image - folder 3. Make sure that the following files exists in the system folder (e.g. Windows\System): MFC42.DLL MSVCRT.DLL' ADVAPI32.DLL CBW32.DLL COMCTL32.DLL OLE32.DLL. OLEAUT32.DLL COMDLG32.DLL MSVCRT.DLL VERSION.DLL OLCH2X32.OCX OLEPRO32.DLL OLCHART.LIC 4. Execute the following commando: C:\Windows\System\regsvr32 c:\Windows\System\olch2x32.ocx 5. Create shortcut to FraunDiff in the Start Menu. 6. Start the application. 7. Calibrate the stepmotor by noting the number of steps used from 0.5 mm to 99.5 (Menu: Advanced\Stepmotor\Run). The stepmotor constant (step/mm) is then found by dividing the number of steps by (99.5-0.5 = 99) mm. This number can be entered into the application by the routine Advanced\Stepmotor\Step Constant. 8. Calibrate the detector position by reading off the position on the scale below the detector. The position (in mm) is registered in the program i the routine Advanced\Detector Position Now the application is ready to use. Good luck! Kraftig Data 2.1.2004